Calibration observations were undertaken on a series of secondary calibrators, which are listed with their SCUBA fluxes in Table 1. The data for the FCF calculations were taken between the 23rd of February and the 14th of March, 2010.
1ex]0pt3.5ex Source | RA(J2000) | DEC(J2000) | 850 m flux / Jy | 450 m flux / Jy | Ref |
1ex]0pt3.5ex HL Tau | 04 31 38.4 | +18 13 59.0 | 2.36 0.24 | 9.9 2.0 | [8] |
1ex]0pt3.5ex CRL 618 | 04 42 53.60 | +36 06 53.7 | 4.7 0.37 | 12.1 2.2 | [8] |
1ex]0pt3.5ex CRL 2688 | 21 02 18.81 | +36 41 37.7 | 6.39 0.51 | 30.9 3.8 | [8] |
1ex]0pt3.5ex IRC+10216 | 09 47 57.38 | +13 16 43.7 | 8.8 1.1 | 17.5 4.5 | [8] |
1ex]0pt3.5ex V883 Ori | 05 38 19 | -07 02 2.0 | 1.34 0.01 | 7.28 0.07 | [9] |
1ex]0pt3.5ex Alpha Ori | 5 55 10.31 | +07 24 25.4 | 0.628 0.008 | 1.39 0.04 | [9] |
1ex]0pt3.5ex TW Hydrae | 11 01 51.91 | -34 42 17.0 | 1.37 0.01 | 3.9 0.7 | [9] |
1ex]0pt3.5ex Arp 220 | 15 34 57.21 | +23 30 09.5 | 0.668 0.007 | 2.77 0.06 | [9] |
The observations were reduced with the mapmaker using the config dimmconfig_bright_compact.lis
and post-processed with the Picard recipe SCUBA2_FCFNEFD
. Figure 29 shows the
850 m and
450 m values
for all calibrator observations taken during the S2SRO period. The resulting mean FCF’s in each
waveband are as follows:
(11) |
(12) |
The first note regarding the FCF’s produced by the current reduction is in regard to the picowatt (pW) scale produced by the mapmaker. The pW scale is dependent on the accuracy of the heater resistance and the fraction of the heater power which is transferred to the bolometer. At the time of data release, the effective resistance was not well described, which induces an uncertainty in the pW scale of the resultant maps. This resistor value, and therefore the absolute pW scale, will be determined accurately when the instrument is returned to operation, and when this is known the new values will be added to the reduction code. This will affect the pW values of all observations, and the corresponding FCFs. However, the flux scaling will be preserved. The FCF values above have been calculated with the original pW scale.
Secondly, it is obvious that there is a large scatter in the FCFs in Figure 29. It is possible that the variation in the FCFs is produced by instrument performance changes or inconsistencies in the way the map-maker reduces the observations. At the time of release, the source of the scatter was not well understood and investigation continues. However, it is worth noting that the scatter between calibrations in an individual night of observations was sometimes observed to be as high as the scatter over the entire dataset. No trends were observed as a function of atmospheric transmission, or time during the night, or over the entire observing period. It is for this reason that it is advised that the average FCF values above are used, as opposed to selecting individual calibrations and using that FCF to calibrate your data.
It is the recommendation to use the mean values presented in the previous section to calibrate your data. However, if it is desired to produce an individual FCF from the night of a particular set of science observations, then the method is described here.
dimmconfig_bright_compact.lis
config file
.
SCUBA2_FCFNEFD
on your reduced calibration observation. This will
produce information to the screen and a logfile log.fcfnefd
with the FCFs as mentioned
above, and an NEFD for the observation. Picard by default uses fixed FCF’s to calculate
the NEFD. (450 m:
400 Jy/beam/pW and 850 m:
500 Jy/beam/pW). If you wish to get an NEFD using the FCF calculated for the individual
calibrator you are reducing, add USEFCF=1
to your parameter file.
Part of the risk in S2SRO was that the instrument was being commissioned in parallel to being used for science observations. During February and March 2010 a number of events occurred that will possibly affect the data quality in a good or a bad way. This section documents these changes to aid in interpreting unexpected results that may come out of the data reduction process.
Note: all dates listed below are UT and are inclusive, and although setup changes should not affect the calibration they will affect the number and quality of functioning bolometers.
Up to 20100219 the first and last file in a scan are dark frames for science, pointing9 and focus observations. From 20100220 to 20100302 the first and last file are fast flatfield ramps. From 20100303 the first file is a dark frame and the second and last file are fast ramps. Note noise and discrete flat fields are different.
Until scan #17 on 20100211 discrete flatfields were reduced using the TABLE
method.
From scan #18 on 20100211 discrete flatfields were reduced using the POLYNOMIAL
meth.
Note: the stand alone flatfield observations are not used after the fast ramp flatfield was
implemented 20100223. The fast flatfields are done as part of the observations.
Until scan #45 the heater step in the discrete flatfield was smaller leading to failure of the flatfield on the sky, particular at 450. Hence a lot of the flatfields on these dates were in the dark. This means less accurate flatfields.
Fast ramp flatfield used implemented but done in the dark. Smurf uses the discrete flatfields for these dates.
Fast ramp flatfield on the sky. Note: on 20100223 the data header claims the ramp was in the dark but from notes and the heater value it was on the sky i.e. the dark shutter header was wrong this night. Smurf knows about this and uses the fast ramp flatfield.
Fast ramp flatfield used with an initial dark in the observations. This dark was used as a sanity check and is not used by the map-maker.
Explanatory Notes:
TABLE
and POLYNOMIAL
are two reduction methods used for discrete flatfields (fast ramp
always uses a polynomial method). Due to the lack of dark frame subtraction the TABLE
flatfield only used 2 points of the 10
heater settings in a discrete flatfield. Further, the slope from these two points were extrapolated
far away. Obviously this affects the flatfield accuracy and quality. The flat field calculation
should be redone using the POLYNOMIAL
method for data taken before 2010021110.
s4a detector bias set to 40000 - during the S2SRO the normal s4a value was 65000.
The heater tracking was changed such that the heater is returned to the default value each time the shutter was closed. This was done to prevents drifts in the heater setting. Such drift affect the setup. However, noise observations also reset the heater. Thus the heater drift before this adjustment were small and is not believed to have affected the data.
A large number of spikes are present in the s4a array data.
The MCE was changed on the s4a array significantly decreasing the number of spikes in the s4a data.
Very bad seeing: data severely affected
For early commissioning data it was found that one of the mirrors was installed upside down leading to a slightly distorted beam shape. This was fixed from 20091203 so care should be taken when analysing very early commissioning data found in the archive before that date.
A two component pointing model (utilising only the CA
and IE
terms) was used, leading
to large pointing shifts when doing large slews. On 20100226 the full eight component
model was implemented and all-sky pointing improved. This should not affect data
quality significantly since local pointing would still be adequate even with the earlier
model.
9Pointings also have dark frames between the data frames in a scan for most of the SRO period.
10Re-reduce the relevant flatfields using calcflat and then use copyflat to copy the flatfield into the data files before using the map-maker.